Vitamin E (Tocopherol)

Message
Serum must be poured off, do not send in mother tube.


Test Code
VITE


Quest Code
931


CPT Codes
84446

Includes
Vitamin E, Alpha-Tocopherol and Vitamin E, Beta-Gamma-Tocopherol


Preferred Specimen
2 mL serum - protected from light


Patient Preparation
Overnight fasting is required

Minimum Volume
0.7 mL


Instructions
Separate from cells as soon as possible after clotting. Aliquot and send serum in an amber tube. If an amber tube is not available, wrap aliquot tube in aluminum foil to protect from light.


Transport Temperature
Refrigerated (cold packs)


Specimen Stability
Room temperature: 24 hours
Refrigerated: 7 days
Frozen: 28 days


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Gross hemolysis • Grossly lipemic sera • Received room temperature • Not light protected • Plasma • Serum unseparated from the clot • Serum in serum separator tube not separated from the gel


Methodology
Chromatography

FDA Status
This test was developed and its analytical performance characteristics have been determined by Quest Diagnostics. It has not been cleared or approved by FDA. This assay has been validated pursuant to the CLIA regulations and is used for clinical purposes.

Setup Schedule
Set up: Mon-Sat; Report Available: 3 days


Reference Range
See Laboratory Report


Clinical Significance
Deficiency of vitamin E may cause extensive neuropathy in young children and, in addition, is suspect as a possible cause of motor and sensory neuropathy in older children and in adults. One likely cause of vitamin E deficiency is intestinal malabsorption, resulting from bowel disease, pancreatic disease, or chronic cholestasis. Other causes of malabsorption of vitamin E include celiac disease, cystic fibrosis, and intestinal lymphangiectasia.


Performing Laboratory
MedFusion
2501 South State Highway 121 Suite 1100
Lewisville, TX 75067-8188



The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.