Immune Cell Function (NY) [17836X]

Test Code
17836


CPT Codes
86352<br><strong>This test code is for New York patient testing. For non-New York patient testing, use test code 15435.</strong>

Preferred Specimen
1 mL whole blood collected in a sodium heparin (green-top) tube
This test is restricted for transplant patients only.


Minimum Volume
0.5 mL


Instructions
Restricted use for transplant facilities only. Contact the Lab before ordering for Special Logistics arrangements.

Sample must be in laboratory by 4:00 p.m. on the day of set up.
Sample must be collected after 6 a.m. if it will not be received by 4 p.m. on the day of set up.

The Immune Cell Function Assay must be setup within 30 hours of collection. The test is set up Tuesday through Friday from 11 a.m. to 6:30 p.m. Samples should be collected Monday to Thursday. Samples received beyond the 30 hours stability will be rejected.


Transport Temperature
Room temperature


Specimen Stability
Room temperature: 30 hours
Refrigerated: Unacceptable
Frozen: Unacceptable


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Gross hemolysis • Clotted • Samples received >30 hours after collection • If specimen viability is <80%, test will not be run


Methodology
Luminescent Detection

Setup Schedule
Set up: Tues-Fri; Report available: 1-2 days


Reference Range
≤225 ng/mL ATP: Low Immune Cell Response
226-524 ng/mL ATP: Moderate Immune Cell Response
≥525 ng/mL ATP: Strong Immune Cell Response


Clinical Significance

This non-pathogen-specific immune function assay detects cell-mediated immunity by measuring intracellular adenosine triphosphate (iATP) levels in stimulated CD4 lymphocytes. This test may be useful in assessing the immunosuppressive status of solid organ transplant (SOT) recipients to help predict the risk of infection and optimize immunosuppressive regimens [1,2].

The risk of infection after transplantation depends upon the epidemiologic exposures and the net state of immunosuppression [3]. Excessive immunosuppression increases the risk of infection; inadequate immunosuppression increases the risk of graft rejection. Because CD4 lymphocytes mediate cell-mediated immunity (CMI), quantification of CD4 activation may be helpful in assessing the immunosuppression status after SOT. Early responses of CD4 lymphocytes can be quantified by measuring iATP production in stimulated CD4 cells. By assessing CMI, this assay may help estimate the aggregate impact of multiple factors, such as the use of immune suppressants and underlying clinical conditions, on a patient's immune status [1,2].

This assay cannot be used as a diagnostic test for infection or graft rejection and does not replace diagnostic assays (eg, culture, viral load, and biopsy) to inform clinical decisions [2]. Without standardized strategies to simultaneously assess a patient's risk for infection and graft rejection, clinical judgment remains crucial for individualizing patient management. The value of this assay for assessing risk of infection [4] and rejection [4,5] in SOT recipients needs further study.

Results are most accurate if testing is performed on the day of blood draw. Performing the assay ≥1 day after the draw may lead to falsely low iATP levels [6].

The results of this test should be interpreted in the context of pertinent clinical and family history and physical examination findings.

References
1. ImmuKnow immune cell function assay. Eurofins Viracor. Accessed March 1, 2022. https://www.eurofins-viracor.com/clinical/our-testing/immuknow/
2. Kowalski RJ, et al. Transplantation. 2006;82(5):663-668.
3. Fishman JA. Am J Transplant. 2017;17(4):856-879.
4. Ling X, et al. Transplantation. 2012;93(7):737-743.
5. Rodrigo E, et al. Liver Transpl. 2012;18(10):1245-1253.
6. Suviolahti E, et al. Transplantation. 2012;94(12):1243-1249.



Performing Laboratory
Quest Diagnostics Nichols Institute
14225 Newbrook Drive
Chantilly, VA 20153



The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.