Iodine, Serum/Plasma [16599]

Message
For Specimen Integrity during Extreme Weather see the “Lockbox Usage in Extreme Weather” document at the top of this page.


Test Code
IODINE


Alias/See Also
16599


CPT Codes
82542

Preferred Specimen
1 mL serum collected in a no additive trace metal-free (dark blue-top) tube
or
1 mL plasma collected in an EDTA trace metal-free (royal blue-top) tube
Transfer to an acid washed plastic vial for shipment


Minimum Volume
0.5 mL


Instructions
Separate plasma or serum from cells and pour off into an acid washed vial (AWV) or other trace element free container.


Transport Container
Acid washed plastic vial


Transport Temperature
Room temperature


Specimen Stability
Room temperature: 10 days
Refrigerated: 10 days
Frozen: Unacceptable


Methodology
Inductively Coupled Plasma/Mass Spectrometry (ICP/MS)

FDA Status
This test was developed and its analytical performance characteristics have been determined by Quest Diagnostics. It has not been cleared or approved by the U.S. Food and Drug Administration. This assay has been validated pursuant to the CLIA regulations and is used for clinical purposes.

Setup Schedule
Set up: Tues-Sat p.m.; Report available: 1-3 days


Reference Range
52-109 mcg/L


Clinical Significance
Iodine is an essential element that is required for thyroid hormone production. The measurement of iodine serves as an index of adequate dietary intake and iodine overload, particularly from iodine-containing drugs such as Amiodarone.


Performing Laboratory
Quest Diagnostics Nichols Institute
14225 Newbrook Drive
Chantilly, VA 20153


Last Updated: November 30, 2021


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.