Resources

LEAD, WHOLE BLOOD

Message
Lavender EDTA vacutainer cannot be used due to lead concentration in the stopper.
The only EDTA accepted is the microtainer.


Test Code
LEAD2


Alias/See Also
Lead


CPT Codes
83655

Preferred Specimen
1 Tan tube (lead-free EDTA)


Minimum Volume
0.5 mL


Other Acceptable Specimens
  • 1 Royal blue tube (EDTA,trace element free -do not collect Royal blue for serum)
  • 2 Lavender microtainers will be accepted


Transport Temperature
Room temperature or refrigerate


Specimen Stability
  • 15-30°C up to 4 months
  • 2-8°C up to 4 months


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
  • Clotted samples will be rejected
  • Samples collected in a EDTA vacutainer tube will be rejected


Methodology
Inductively coupled plasma mass spectrometry

Setup Schedule
M, Tu, W, Th, F


Report Available
Report available within 24 hours of receipt at the performing laboratory.


Reference Range
Age Value (ug/dL)
0-16 years 0 - 3.4
Adult 0 - 4.9

 


Clinical Significance
Recommended for routine testing for lead exposure.


Performed By
Alverno Laboratories

Performing Laboratory
Alverno Central Lab  
NCICP (Chemistry, ICP-MS)


Last Updated: June 14, 2023


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.