GLUCOSE, NON FASTING

Test Code
LAB304128


CPT Codes
82947

Preferred Specimen
One 4 mL gray top tube
Pediatric:  One gray top microtainer


Other Acceptable Specimens
Light green top, dark green top, red top or gold top tube


Instructions
Sample must be centrifuged within 2 hours of collection or within 4 hours if drawn in a Gray Top Tube. Samples collected in a red (Serum), dark green (heparin), or Gray (sodium fluoride) top tube must be removed from the red cells and put into aliquot tubes following centrifugation.


Transport Temperature
Refrigerated


Specimen Stability
Unstable, centrifuge within 2 hours of collection (Red/ Dark Green/ Light Green/ Gold) or within 4 hours (Gray)

Gray Top tube:  
Ambient: 2 days  
Refrigerated: 7 days  
Frozen: 3 months

Red/Dark Green Top tube:  
Ambient: 2 hours  
Refrigerated: Not specified  
Frozen: Not specified

Light Green/ Gold Top tube:  
Ambient: 8 hours  
Refrigerated: 72 hours  
Frozen: Not specified


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Gross hemolysis


Methodology
Glucose Oxidase

Setup Schedule
Sunday-Saturday


Reference Range
Therapeutic Range: 65-125 mg/dL  

Critical level:
0-7 days: <=40 or >=200 mg/dL
8-90 days: <=40 or >=300 mg/dL
>=91 days: <=53 or >=400 mg/dL


Performing Laboratory
West Virginia University Hospital, Inc.
Bluefield Pavilion Campus
Berkeley Medical Center
Barnesville Hospital
Braxton County Memorial
Camden Clark Medical Center
Camden Clark Medical Center
Fairmont Medical Center
Garrett Regional Medical Center
Harrison Community Hospital
Jackson General Hospital
Jefferson Medical Center
Princeton Community Hospital
Potomac Valley Hospital
Reynolds Memorial Hospital
Summersville Regional Medical Center
St. Joseph’s Hospital
United Hospital Center
Uniontown Hospital
Wetzel County Hospital
Wheeling Hospital




The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.