Selenium (SELEN)

Message
Patient should refrain from taking vitamins or mineral supplements at least three days prior to specimen collection.


Test Code
LAB579


Quest Code
5507


CPT Codes
84255

Preferred Specimen
2 mL serum collected in a no additive (royal blue-top) tube in a trace-element package


Patient Preparation
Patient should refrain from taking vitamins or mineral supplements at least three days prior to specimen collection

Minimum Volume
0.7 mL


Other Acceptable Specimens
Plasma collected in: EDTA (royal blue-top) tube or sodium heparin (royal blue-top) tube


Instructions
Carefully clean skin prior to venipuncture. Avoid worksite collection. Avoid hemolysis. Use the royal blue-top "Trace Metal" evacuated tube, which is available with EDTA, heparin or no additives. Do not use gel barrier tubes. Centrifuge serum or plasma specimens within 1 hour of collection. Immediately separate serum and plasma specimens from the cells into trace element collection vial(s).


Transport Container
No additive (royal blue-top) tube in trace-element package


Transport Temperature
Refrigerated (cold packs)


Specimen Stability
Room temperature: 8 hours
Refrigerated: 14 days
Frozen: 30 days


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Hemolysis • Received room temperature • Serum separator tube (SST)


Methodology
Inductively Coupled Plasma/Mass Spectrometry (ICP/MS)

Setup Schedule
Tuesday, friday


Reference Range
<2 Years 16-71 mcg/L
2-4 Years 40-103 mcg/L
4-16 Years 55-134 mcg/L
Adult 63-160 mcg/L


Clinical Significance
Selenium is an element of parenteral nutrition. Monitoring the Selenium concentration is useful in assessing parenteral nutrition, especially recent intake. Concentrations are also monitored in children with propionic acidemia who require special diets with supplements.


Performing Laboratory
Quest Diagnostics Nichols Institute-Chantilly VA
14225 Newbrook Drive
Chantilly, VA 20151-2228




The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.