| A B C D E F G H I J K L M N O P Q R S T U V W X Y Z # |
Cerner: TEM-Profile Extrinsic Pathway, TEM-Profile Instrinsic Pathway, and TEM-Profile Fibrinogen Extrinsic Pathway.
Softlab: ROTEM, ROTMM, and RTIME
MessageViscoelastic testing can be useful in multiple settings where there is a need to measure the clotting process, beginning with fibrin formation through the lysis of the clot. The assays analyze the viscoeleastic properties of whole blood samples under low-shear conditions and display results both graphically and numerically. One of the avalible platforms we have is Rotational thromboelastometry (ROTEM).
ROTEM is proven effective in managing coagulation when multiple factor deficiencies are involved, as for perioperative and trauma patients who often require transfusion of multiple components and factor concentrates.
Contact Pathologist on-call to help interpret results.
ROTEM is proven effective in managing coagulation when multiple factor deficiencies are involved, as for perioperative and trauma patients who often require transfusion of multiple components and factor concentrates.
Contact Pathologist on-call to help interpret results.
Test Code
ROTEM
Alias/See Also
Rotational thromboelastometry
Minimum Volume
3mL light blue tube (sodium citrate) filled to the indicated fill line.

Instructions
Draw 2mL discard before collecting sample Fill tube to the line or arrow.
Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Clotted
If the tube is not filled up to the line, or if tube is overfilled.
If the tube is not filled up to the line, or if tube is overfilled.
Performing Laboratory
UMC Regional Laboratory