Beta-2-Glycoprotein I Antibodies (IgG, IgA, IgM)

Message
*Take to processing to be spun and frozen ASAP.


Test Code
30340


Alias/See Also
LAB1179


CPT Codes
86146 (x3)

Preferred Specimen

3 mL plasma collected in a 3.2% sodium citrate (light blue-top) tube



Minimum Volume
1.5 mL


Other Acceptable Specimens
Serum


Instructions
Plasma: Centrifuge light blue-top tube 15 minutes at approximately 1500 g within 60 minutes of collection. Using a plastic pipette, remove plasma, taking care to avoid the WBC/platelet buffy layer and place into a plastic vial. Centrifuge a second time and transfer platelet-poor plasma into a new plastic vial. Plasma must be free of platelets (<10,000/mcL).


Transport Container
Transport tube


Transport Temperature

Room temperature



Specimen Stability

Room temperature: 5 days
Refrigerated: 14 days
Frozen: 30 days



Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Gross hemolysis • Grossly lipemic


Methodology
Immunoassay (IA)

Limitations
Factor VIII inhibitors may cause a false-positive lupus anticoagulation test. The temporary presence of antiphospholipid antibodies may accompany infections or certain drugs.


Reference Range
B2-Glycoprotein I (IgG) Ab <20 U/mL Antibody not detected
≥20 U/mL Antibody detected
B2-Glycoprotein I (IgA) Ab <20 U/mL Antibody not detected
≥20 U/mL Antibody detected
B2-Glycoprotein I (IgM) Ab <20 U/mL Antibody not detected
≥20 U/mL Antibody detected


Clinical Significance
Beta-2-Glycoprotein 1 (also known as apolipoprotein H), is a cofactor in antiphospholipid antibody binding and is the critical antigen in the antiphospholipid antibody syndrome. The presence of high titer IgG/IgM beta-2-glycoprotein 1 antibodies is one criteria for the diagnosis of antiphospholipid antibody syndrome.


Performing Laboratory
Quest Diagnostics Nichols Institute-San Juan Capistrano, CA
33608 Ortega Highway
San Juan Capistrano, CA 92675-2042




The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.