Fibrinogen

Test Code
FIB


CPT Codes
85384

Preferred Specimen

1 Sodium Citrate Plasma - Blue



Minimum Volume
Blue tops must be filled to Minimum Fill Indicator Line on tube.


Other Acceptable Specimens
none


Instructions

Outreach clients should double-spin the blue top plasma, remove the plasma from the RBC button, and freeze in a polypropylene tube.



Transport Container

Blue, Polypropylene tube



Transport Temperature
Blue - RT Polypropylene tube - Frozen


Specimen Stability
Whole Blood - 4 hours RT Separated plasma - 8 hours at RT or 2 weeks frozen


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Sample must be free of clots and hemolysis.


Methodology
Electromagnetic Viscosity Detection

Setup Schedule

Daily on Receipt



Report Available
Upon Completion of Analysis


Reference Range

200 - 424 mg/dL



Clinical Significance

This test is intended for the quantitative determination of fibrinogen in plasma by the Clauss clotting method. In the presence of an excess of thrombin, the clotting time of diluted plasma is inversely proportional to the level of plasma fibrinogen. Increases of fibrinogeen are observed in cases of diabetes, inflammatory syndromes and obesity. A decrease of the fibrinogen level is observed in DIC, fibrinolysis and hereditary diseases.





The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.