Chlamydia trachomatis, TMA (Alternate Target), Urogenital

Test Code
15031


CPT Codes
87491

Preferred Specimen
1 vaginal, endocervical or urethral swab collected in an Aptima® Vaginal Swab Specimen Collection kit or Multitest collection kit, or Unisex swab specimen collection kit or
2 mL urine with no preservative transferred into an Aptima® urine collection kit or
1 mL PreservCyt® (ThinPrep®) transferred into Aptima® Specimen Transfer tube (green label) or Aptima® Vaginal Swab Specimen Collection kit or Multitest collection kit Collection tube (orange label)


Instructions
Swabs: Swab must be submitted in Aptima® Multitest Collection Kit (orange label), or Unisex Swab Specimen Collection Tube (white label). Follow instructions in the Aptima® Specimen Collection Kits for Vaginal, Endocervical and Urethral Swab Specimens package insert. In females, to ensure collection of an adequate specimen, columnar epithelial cells lining the endocervix should be obtained. To that effect, excess mucus should be removed prior to sampling by using the white shaft cleaning swab which is discarded after use.
Collection Container: Aptima® Multitest Collection Kit (orange label), or Unisex Swab Specimen Collection Kit (white label).

Urine: Urine must be submitted in Aptima® Urine Specimen Collection Tube within 24 hours of collection. The patient should not have urinated for at least one hour prior to specimen collection. Direct patient to provide a first-catch urine (approximately 20-30 mL of the initial urine stream) into a urine collection cup free of any preservatives. Collection of larger volumes of urine may result in specimen dilution that may reduce test sensitivity. Female patients should not cleanse the labial area prior to providing the specimen. Thus, a urine sample collected from a female for Chlamydia/GC TMA cannot be collected at the same voiding event as that urine which would be intended for bacterial culture. Remove the cap of the Aptima® Urine Specimen Collection Tube and transfer 2 mL of urine into the urine specimen transport tube using the disposable pipette provided. The correct volume of urine has been added when the fluid level is between the black lines within the clear pane on the urine specimen transport tube label.
Collection Container: Aptima® Urine Specimen Collection Tube.

PreservCyt® (ThinPrep®): Transfer 1 mL of ThinPrep PreservCyt solution into Aptima® Specimen Transfer Tube (green label) or Aptima® Multitest Collection Tube (orange label).


Transport Container
Aptima® transport tube


Transport Temperature
Room temperature


Specimen Stability
Endocervical, urethral and vaginal swabs
Room temperature: 60 days
Refrigerated: 60 days
Frozen: 1 year

Urine in Aptima® urine collection kit
Room temperature: 30 days
Refrigerated: 30 days
Frozen: 1 year

PreservCyt® solution specimen in Aptima®
Room temperature: 14 days
Refrigerated: 30 days
Frozen: 1 year


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Transport tubes with 2 swabs • Non-Aptima® transport tubes and swabs • Transport tubes containing white cleaning swabs • Over or underfilled Aptima® urine tubes


Methodology
Dual Kinetic Assay (DKA) • Target Capture • Transcription-Mediated Amplification (TMA)

Setup Schedule
Tues


Report Available
3-11 days


Reference Range
Not detected


Clinical Significance
Several methods are available for the detection of C. trachomatis in clinical specimens, including giemsa staining of tissue, fluorescent antibody staining, culture, and nucleic acid detection. Although culture is very specific, it lacks sensitivity, so recommendations have been put forth to include multiple tests in assisting with a diagnosis of infection, including tests such as the CT/NG Aptima TMA Assay.


Performing Laboratory
Quest Diagnostics Nichols Institute-Chantilly VA
14225 Newbrook Drive
Chantilly, VA 20151-2228



Last Updated: November 6, 2024


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.