Post-Transplant/Immunocompromised, Adenovirus Quant PCR (LABCORP)

Test Code
1230120206


Alias/See Also
LAB2210: Post-Transplant/Immunocompromised, Adenovirus Quant PCR | LABCORP EAP: 815931


CPT Codes
87799x1

Preferred Specimen
LABCORP LAB: LAVENDER TOP-EDTA (Frozen) Min Vol: .5


Other Acceptable Specimens
LABCORP LAB: PALE YELLOW (ACD): (Frozen) Min Vol: .5


Instructions
Trasylol(R) kits may be ordered through the PeopleSoft system (LabCorp #33328). Using a chilled 6-mL lavender-top (EDTA) tube taken from the kit, collect a whole blood specimen. Mix the specimen several times by inverting the EDTA collection tube. After removing the cap from the EDTA draw tube, take one of the sterile, Beral pipettes (from under the grey foam), and add 0.25 mL Trasylol(R) to the EDTA tube. Recap the EDTA tube and invert several times to mix well. Centrifuge the EDTA tube to separate the plasma from the cells, and immediately transfer the plasma into one of the brown screw-cap transfer tubes provided in the kit. There should be a ""Trasylol(R) Added"" label affixed to the brown transfer tubes. Cap and freeze the labeled transfer tube containing the EDTA plasma with Trasylol(R) added. To avoid delays in turnaround time when requesting multiple tests on frozen samples, PLEASE SUBMIT SEPARATE FROZEN SPECIMEN FOR EACH TEST REQUESTED


Transport Temperature
Stability Room Temp: Unstable Refrigerated: 1 day Frozen: 14 days Cause for Rejection Sample not collected with Trasylol(R); sample not submitted in tube with Trasylol label; gross hemolysis; recently administered radioisotopes; specimen not received frozen; serum, sodium citrate, or heparinized plasma specimen; lipemia


Report Available
LABCORP LAB
STAT: 5760 min
Routine: 5760 min
Extended TAT: No TAT min
Timed: 5760 min
Life or Death: 5760 min



Last Updated: March 6, 2023


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.