Protein S, Antigen (LABCORP)

Test Code
99901000579


Alias/See Also
LAB760: Protein S Antigen, Free and Total | LABCORP EAP: 164517


CPT Codes
85306x1, 85305x1

Preferred Specimen
LABCORP LAB: LT BLUE TOP-CITRATE (Frozen) Min Vol: 1


Instructions
Preferred specimen: Plasma collected in Light blue top tube (sodium citrate)


Transport Temperature
NOTE: Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results. Specimen: Plasma, frozen Volume: 1 mL Transport Container: Plastic vial containing plasma from Light Blue top tube Specimen Collection: 1. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again. 2. Aliquot plasma into a plastic vial leaving 0.25 mL in the bottom of centrifuged vial. 3. Freeze plasma immediately (no longer than 4 hours after collection) at -20 degrees C or, ideally, at < or =-40 degrees Transport Temperature: Frozen Turn-around-time: 2-3 days Cause for Rejection: Gross hemolysis; clotted specimen; frozen specimen thawed in transit; improper labeling.


Report Available
LABCORP LAB
STAT: 4320 min
Routine: 4320 min
Extended TAT: No TAT min
Timed: 4320 min
Life or Death: 4320 min



Last Updated: March 6, 2023


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.