Cardio IQ® Homocysteine

Test Code
91733Q


Quest Code
91733


CPT Codes
83090

Preferred Specimen
1 mL serum


Minimum Volume
0.5 mL


Instructions
Red-top tubes (no gel): Place the specimen in a refrigerator or ice bath for 30 minutes after collection. Centrifuge the specimen as soon as possible after complete clot formation has taken place. Transfer the serum to a transport tube.

Barrier gel separator tubes: Place the specimen in a refrigerator for 30 minutes after collection. Centrifuge the specimen as soon as possible after complete clot formation has taken place. Do not place barrier tubes in an ice bath as freezing may prevent the barrier gel from adequately separating serum from cells. Transfer serum to a transport tube.


Transport Container
Transport tube


Transport Temperature
Refrigerated (cold packs)


Specimen Stability
Room temperature: 4 days
Refrigerated: 14 days
Frozen: 6 months


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Gross hemolysis • Grossly lipemic • Unseparated serum or plasma • Whole blood specimens


Methodology
Enzymatic

Setup Schedule
Mon-Fri


Reference Range
AgeMale 
(umol/L)
Female 
(umol/L)
0-1 yearsNot EstablishedNot Established
2-9 years≤7.6≤7.6
10-14 years≤9.7≤8.4
15-17 years≤11.3≤10.2
18-29 years≤12.9≤10.9
30-49 years≤13.5≤11.0
>49 years≤15.2≤13.4


Clinical Significance
An elevated concentration of Homocysteine is an independent risk factor for cardiovascular disease.


Performing Laboratory
Cleveland HeartLab, Inc
6701 Carnegie Avenue, Suite 500
Cleveland, OH 44103-4623




The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.