A B C D E F G H I J K L M N O P Q R S T U V W X Y Z # |
Bilirubin Direct
Test CodeBILID
Preferred Specimen
1 mL Serum or Plasma (lithium heparin) or EDTA
Instructions
"Separate from cells ASAP with a max of 2 hours from collection time. Bilirubin is photosensitive, protect from daylight and fluorescent light
Testing Frequency: Performed as ordered"
Testing Frequency: Performed as ordered"
Transport Container
PST, Plasma Separator
Transport Temperature
Refrigerate
Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Stability: 5 days Refrigerated
Methodology
Jendrassik-Grof w/blank
Reference Range
"M Up to 30 Days 0 - 0.3 mg/dL
M Up to 999 Years 0 - 0.2 mg/dL
F Up to 30 Days 0 - 0.3 mg/dL
F Up to 999 Years 0 - 0.2 mg/dL"
M Up to 999 Years 0 - 0.2 mg/dL
F Up to 30 Days 0 - 0.3 mg/dL
F Up to 999 Years 0 - 0.2 mg/dL"