Dialysis Water Culture : 1000266

Test Code
DIWC or 1000266


Alias/See Also
Dialysate Culture


CPT Codes
87070

Instructions
Dialysis Water:
1. Dialysate should be collected post-dialyzer by disconnecting line and allowing fluid to flow for at least 30-60 seconds before collecting in a sterile container or endotoxin free container.
2. Product water from the distribution system should be collected from the spigot at the point before the water goes to the treatment area. The water should run for 1 minute before collected.


Transport Container
See specimen collection information. Submit 20 mL dialysis water.


Transport Temperature
Refrigerated.


Specimen Stability
Transport as quickly as possible.


Methodology
Aerobic culture following AAMI standards

Setup Schedule
Sunday - Saturday


Report Available
7 days


Reference Range
Maximum acceptable count <200 colonies/mL, with an action limit of 50 colonies /mL


Performing Laboratory
med fusion



The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.