TSH with HAMA Treatment

Test Code
TSHHM


Alias/See Also

LAB11916



CPT Codes
<p>84443 (x2)</p>

Includes
Ultrasensitive TSH with and without HAMA treatment


Preferred Specimen

Patient Preparation: Specimen collection after fluorescein dye angiography should be delayed for at least 3 days. For patients on hemodialysis, specimen collection should be delayed for 2 weeks. According to the assay manufacturer Siemens: "Samples containing fluorescein can produce falsely depressed values when tested with the Advia Centaur TSH3 Ultra assay. "


Specimen Type: Serum


Collection Container/Type


Preferred: Serum gel


Acceptable: Red top


Submission Container/Tube: Plastic vial


Specimen Volume: 2 mL



Patient Preparation
Specimen collection after fluorescein dye angiography should be delayed for at least 3 days. For patients on hemodialysis, specimen collection should be delayed for 2 weeks. According to the assay manufacturer Siemens: "Samples containing fluorescein can produce falsely depressed values when tested with the Advia Centaur TSH3 Ultra Assay."

Minimum Volume

1.5 mL



Transport Container
Plastic screw-cap vial


Transport Temperature
Room temperature


Specimen Stability

Room temperature: 7 days

Refrigerated: 7 days

Frozen: 28 days



Methodology

Immunoassay (IA)



Setup Schedule
Set up: Tues-Sat; Report available: 2-3 days


Report Available

2-3 days



Reference Range
TSH, HAMA Treated See below
TSH, Untreated  
Premature infant (28-36 weeks), 1st week of life 0.20-27.90 mIU/L
Term infant (>37 Weeks), serum or cord blood  1.00-39.00 mIU/L
1-2 Days 3.20-34.60 mIU/L
3-4 Days 0.7-15.40 mIU/L
5 Days-4 Weeks 1.70-9.10 mIU/L
1-11 Months  0.80-8.20 mIU/L
1-19 Years 0.50-4.30 mIU/L
≥20 Years 0.40-4.50 mIU/L
Pregnancy  
  First Trimester 0.26-2.66 mIU/L
  Second Trimester  0.55-2.73 mIU/L
  Third Trimester 0.43-2.91 mIU/L


Clinical Significance
Precision in the estimation of TSH is crucial to the follow-up of subjects with thyroid disorders. The preincubation of samples to remove HAMA immunoglobulins provides a way to ensure that an otherwise unsuspected rise or drop in TSH concentration will not be interpreted as either a false positive or false negative.


Performing Laboratory

Quest Diagnostics Nichols Inst San Juan Capistrano
33608 Ortega Highway
San Juan Capistrano, CA 92675




Last Updated: February 17, 2023


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.