GLUCOSE, RANDOM

Message
Chemistry


Test Code
LAB82


CPT Codes
82947

Preferred Specimen
Light green top tube (lithium heparin plasma).
Fasting specimen preferred.


Minimum Volume
Whole blood: 600 uL (one full light-green-top microtainer or two full red-top microtainers).

Serum/plasma: 300 uL plasma.


Other Acceptable Specimens
Gold top (serum separator tube) or Red Top
Gray Top tube acceptable


Instructions
Fasting 8 - 12 hours recommended Included in Basic Metabolic Panel and Comprehensive Metabolic Panel. LAB: Centrifuge and analyze promptly. Serum acceptable. Serum or plasma: Hemolysis decreases results. Dilute grossly lipemic specimens. Separate specimens within 1 hour of collection.


Specimen Stability
Ambient: up to 24 hours.
Refrigerated: 7 days.
Frozen: 1 year.


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Grossly Hemolyzed, contaminated, not enough volume



FDA Status
Approved

Setup Schedule
Everyday


Report Available
Same day


Limitations
In fresh specimens, catalase released from the lysis of red blood cells causes a negative bias in glucose results. The degree of bias is proportional to the degree of hemolysis. In fresh samples, a negative bias of up to 10% may be observed with a level of hemolysis associated with a hemoglobin concentration of 250 mg/dL.


Reference Range
65 - 99 mg/dL


Performing Laboratory
Hoag Newport/ Hoag Irvine



The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.