Immunoglobulin M

Message
Chemistry


Test Code
LAB72


CPT Codes
82784

Preferred Specimen
Gold top (serum separator tube)


Minimum Volume
Whole blood: 600 uL (TWO full red-top microtainers or one full light green-top microtainer).

Serum/plasma: 300 uL


Other Acceptable Specimens
Red Top and light green top (lithium heparin plasma) acceptable


Specimen Stability
Ambient: up to 24 hours
Refrigerated: 7 days
Frozen up to 4 weeks


Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Turbid Specimens.


Setup Schedule
Everyday


Report Available
Same day


Limitations
An abnormally elevated level of a monoclonal immunoglobulin, as in cases of multiple myeloma, amyloidosis, or other lymphoproliferative diseases, may interfere with the quantification of IgM.


Reference Range
40 - 230 mg/dL


Performing Laboratory
Hoag Newport



The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.