Factor VIII, Activity

Message
Coagulation


Test Code
LAB306


Alias/See Also
FVIII F8 Factor 8
Factor 8 Assay
 


CPT Codes
85240

Preferred Specimen
Blue top 3.2% citrated tube, plasma 
Note: Do NOT place blue top tube on ice


Minimum Volume
2.7 mL whole blood filled to the designated line on specimen container (blue top tube, citrate)


Other Acceptable Specimens
1.8 mL Low Volume blue top tube, citrated (microtainer)


Instructions
LAB: Centrifuge in validated Platelet Poor Plasma Centrifuge for 2 minutes at 7200 rpm. Specimen to be delivered to Coagulation department STAT. Notify CLS upon arrival.
**Testing must be completed within 4 hours of collection.


Note: If unable to deliver whole blood to testing laboratory within stability time, platelet poor plasma should be aliquoted and placed in freezer. Centrifuge light blue-top for 2 minutes at 7200 rpm (STAT Coag centrifuge) within 60 minutes of collection. Using a plastic pipette, remove plasma, taking care to avoid the WBC/platelet buffy layer and place into a plastic vial. Centrifuge aliquoted plasma a second time and transfer platelet-poor plasma into a new plastic vial and freeze at -20℃ if delay longer than 1 hour. Once thawed, specimen must be tested within 2 hours.

Freeze aliquot at -20℃ for at least 1 hour when sending to alternate campus for testing.


Transport Temperature
Ambient: whole blood (stable 2 hours)
Frozen (-20℃): plasma frozen for at least 1 – 2 hours


Specimen Stability

Whole blood sample stability 
Ambient: 2 hrs
Centrifuged and separated plasma (see instructions) 
Refrigerated 2-8oC: 4 hours
Frozen: -20oC: 2 weeks
Frozen: -70oC: 6 months




Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Hemolyzed, clotted specimen, contaminated, not enough volume, overfilled tube, whole blood specimen transported on ice


Report Available
Same Day or within 24 hours


Reference Range
50 - 150%


Performing Laboratory
HHN



The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.