SARS-CoV2/FLUAB/RSV
MessageMust be collected at Main Hospital or Front Desk Lab ONLY
Test Code
LAB6638
Alias/See Also
COVID 19, COVID, XPRESS SARS-COV-2, SARS COV 2, COV 2
CPT Codes
87637
Includes
SARS-COV-2 RNA by PCR
Influenza A by PCR
Influenza B by PCR
RSV by PCR
Influenza A by PCR
Influenza B by PCR
RSV by PCR
Preferred Specimen
Nasopharyngeal Swab in Viral Transport media
or
Nasal swab in Viral Transport media
or
Nasal swab in Viral Transport media
Minimum Volume
300 uL
Specimen Stability
8 hours room temperature
7 days refrigerated
7 days refrigerated
Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Exceded specimen stability
Incorrect source or collection
Incorrect source or collection
Methodology
Nucleic Acid Amplification PCR
Setup Schedule
As needed
Report Available
65 minutes
Limitations
1. As with any molecular test, mutations within the target regions of the Xpert Xpress SARS-CoV-2/Flu/RSV test could affect primer and/or probe binding resulting in failure to detect the presence of virus or the virus being detected less predictably.
2. Erroneous test results might occur from improper specimen collection; failure to follow the recommended sample collection, handling, and storage procedures.
3. False negative results may occur if virus is present at levels below the analytical limit of detection
4. Viral nucleic acid may persist in vivo, independent of virus viability. Detection of analyte target(s) does not imply that the corresponding virus(es) are infectious or are the causative agents for clinical symptoms.
4. Viral nucleic acid may persist in vivo, independent of virus viability. Detection of analyte target(s) does not imply that the corresponding virus(es) are infectious or are the causative agents for clinical symptoms.
5. This test is a qualitative test and does not provide the quantitative value of detected organism present
6. This test has not been evaluated for monitoring treatment of infection
7. Recent patient exposure to FluMist® or other live attenuated influenza vaccines may cause inaccurate positive results.
Reference Range
Negative
Performing Laboratory
CH CHOMP HOSPITAL LAB (831-625-4811): CH MICROBIOLOGY