A B C D E F G H I J K L M N O P Q R S T U V W X Y Z # |
UR Drugs of Abuse 7 NC
MessageThe screening results are to be used for medical purposes only.
This is a screening test only. Please notify the laboratory within 24 hours if confirmatory testing is required"
This is a screening test only. Please notify the laboratory within 24 hours if confirmatory testing is required"
Test Code
LAB6191
Alias/See Also
DA7NC
Includes
Amphetamines screen – detects d-amphetamine, d-methamphetamine. The assay also detects l‑amphetamine, l‑methamphetamine, d,l-amphetamine, d,l-methamphetamine, methylenedioxyamphetamine (MDA), and, to a lesser extent, methylenedioxymethamphetamine (MDMA) and methylenedioxyethylamine (MDEA).
Barbiturates screen - detects both long-acting (phenobarbital) and short-acting (secobarbital, pentobarbital) barbiturates
Benzodiazepines screen - detects most benzodiazepines and their metabolites
Cannabinoids screen - detects 11-nor- Δ9 –THC-9-carboxylic acid and other Δ9 –THC metabolites
Cocaine screen – detects benzoylecgonine (cocaine metabolite)
Opiates Screen - detects morphine, morphine-3-glucuronide, and codeine in human urine. Also detects synthetic opiates such as hydromorphone, hydrocodone, oxymorphone, oxycodone, dihydrocodeine, meperidine, and levorphanol and the narcotic antagonist nalorphine
PCP screen - detects phencyclidine
Barbiturates screen - detects both long-acting (phenobarbital) and short-acting (secobarbital, pentobarbital) barbiturates
Benzodiazepines screen - detects most benzodiazepines and their metabolites
Cannabinoids screen - detects 11-nor- Δ9 –THC-9-carboxylic acid and other Δ9 –THC metabolites
Cocaine screen – detects benzoylecgonine (cocaine metabolite)
Opiates Screen - detects morphine, morphine-3-glucuronide, and codeine in human urine. Also detects synthetic opiates such as hydromorphone, hydrocodone, oxymorphone, oxycodone, dihydrocodeine, meperidine, and levorphanol and the narcotic antagonist nalorphine
PCP screen - detects phencyclidine
Minimum Volume
1 ml
Instructions
Collect random urine in a clean plastic screw top container without preservative.
Specimen Stability
Refrigerate within 2 hours. Stable for 3 days refrigerated.
Methodology
Two point rate:The concentration of drug in the urine sample is directly propotional to measured enzyme activity.
Limitations
The opiates reagent has cross-reactivity with the semi-synthetic opioids including
oxycodone at 1979 ng/mL and hydrocodone at 250 ng/mL
oxycodone at 1979 ng/mL and hydrocodone at 250 ng/mL
Reference Range
Negative
A negative result indicates that the drug level is below the cut-off concentration:
Amphetamines 1000 ng/mL
Barbiturates 200 ng/mL
Benzodiazepines 200 ng/mL
Cannabinoids 20 ng/mL
Cocaine 300 ng/mL
Opiates 300 ng/mL
Phencyclidine 25 ng/mL
A negative result indicates that the drug level is below the cut-off concentration:
Amphetamines 1000 ng/mL
Barbiturates 200 ng/mL
Benzodiazepines 200 ng/mL
Cannabinoids 20 ng/mL
Cocaine 300 ng/mL
Opiates 300 ng/mL
Phencyclidine 25 ng/mL