A B C D E F G H I J K L M N O P Q R S T U V W X Y Z # |
Platelet Function Analysis
MessageImportant Note
Sample must be collected at Inova Fairfax Medical Campus (outpatient or inpatient).
Test Code
PFA
Preferred Specimen
Patient Preparation:
Specimen Type: Whole Blood Sodium Citrate
Collection Container/Type: 2x Light-blue (citrate)
Submission Container/Tube: Submit sample in original collection container.
Specimen Volume: Two light-blue (citrate) 2.7 mL tubes, completely filled to obtain a 9:1 ratio.
Collection Instructions:
- Venipuncture should be performed using a 21-gauge or larger needle.
- Blood should be drawn directly into an evacuated plastic or siliconized glass tube or syringe containing 3.2% (0.1 05M) buffered sodium citrate (1 part anticoagulant to 9 parts blood).
- After sample collection, ensure proper mixing of anticoagulant by gently inverting the tube by hand 3 to 4 times.
- Discard the sample if there is a venous collapse or stoppage of blood for collection.
- The collection method (both citrate concentration and venipuncture method) should be kept consistent. Use of unbuffered sodium citrate anticoagulant is not recommended.
- Do not refrigerate samples.
- Samples must be walked down directly to the lab. Do not use the tube station.
Additional Information:
Specimen Stability
Room temperature: 4 hours
Refrigerated: Unacceptable
Frozen: Unacceptable
Performing Laboratory
Inova Fairfax Medical Campus Laboratory
3300 Gallows Rd.
Falls Church, VA 22042
Last Updated: December 20, 2023