A B C D E F G H I J K L M N O P Q R S T U V W X Y Z # |
Hemoglobin A1c plus AEG
Test CodeAlias/See Also
A1c
Glycohemoglobin
HBA1C
Hemoglobin A1c
Includes
Average Estimated Glucose
Preferred Specimen
Specimen Type: Whole blood EDTA
Collection Container: Lavender top (EDTA)
Specimen Volume: 3 mL
Minimum Volume
1.0 mL
Instructions
- Be sure to fill tube completely.
- Send specimen in original tube. Do not aliquot.
Specimen Stability
Room temperature: 8 hours
Refrigerated: 7 days
Frozen: Unacceptable
Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)
Unlabeled, mislabeled, wrong tube type, hemolyzed, visually lipemic, clotted, QNS, exceeds specimen/stability requirements.
Methodology
Enzymatic
Setup Schedule
Daily
Report Available
Same day
Limitations
The Hemoglobin A1c assay is susceptible to interference effects from conjugated bilirubin at > 15.0 mg/dL (180 μmol/L),unconjugated bilirubin at > 10.0 mg/dL (171 μmol/L) and Triglycerides >3000 mg/dL.
Glycated HbF is not detected by the assay as it does not contain the β-chain that characterizes HbA1c. However, HbF is measured in the total Hb assay and as a consequence, specimens containing high amounts of HbF (> 5%) may result in lower than expected mmol/mol HbA1c values (IFCC) and %HbA1c values (NGSP).
If patient has a known homozygous or double heterozygous hemoglobinopathy, HBA1C results may be inaccurate. Fructosamine is the method of choice for those patients.
Reference Range
Per ADA guidelines Hemoglobin A1c values of 5.7-6.4%
indicate an increased risk for developing diabetes mellitus.
Hemoglobin A1c values greater than or equal to 6.5% are
diagnostic of diabetes mellitus. A Triglyceride result
greater than 3000 mg/dL can falsely decrease HBA1C results.
Congenital hemoglobinopathy can falsely lower the Hemoglobin
A1c values – Boronate-affinity HbA1c is recommended.
Clinical Significance
Performing Laboratory
Inova Laboratories
2832 Juniper Street
Fairfax, VA 22031
Last Updated: April 19, 2023
Last Review: N. Wolford, April 19, 2023