High Sensitivity Troponin-I

Test Code
HTROP


Alias/See Also
Epic: LAB16407


Preferred Specimen

Specimen Type: Plasma
Collection Container: Green top PPT (Lithium heparin)
Specimen Volume: 3 mL




Minimum Volume

0.25 mL




Other Acceptable Specimens

Specimen Type: Plasma
Collection Container: Green top (Lithium heparin)




Transport Container
Plastic vial


Specimen Stability

Room temperature: 8 hours
Refrigerated: Unacceptable
Frozen: Unacceptable




Reject Criteria (Eg, hemolysis? Lipemia? Thaw/Other?)

Unlabeled, mislabeled, wrong tube type, severely hemolyzed, QNS.




Methodology

Chemiluminescent Microparticle Immunoassay (CMIA)



Setup Schedule
Daily


Report Available

Same day




Limitations

The Alinity i STAT High Sensitivity Troponin-I assay is susceptible to interference effects from total protein > 8.8 g/dL. Total protein from 9.0 to 12.0 g/dL decreased troponin values at 500 ng/L by up to -16.3%.




Reference Range
Male 21-73 years: 0.0-35.0 ng/L
Female 21-75 years: 0.0-14.0 ng/L



Performing Laboratory

Inova Laboratories

2832 Juniper Street

Fairfax, VA 22031



Additional Information
High-Sensitivity Cardiac Troponin I Information & Ordering

Last Updated: May 10, 2023
Last Review: N. Wolford, May 10, 2023


The CPT Codes provided in this document are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payor being billed. Any Profile/panel component may be ordered separately. Reflex tests are performed at an additional charge.